Model |
FLS-1000 |
SX-700*2 |
SX-500*2 |
SX-300*2 |
SX-700BH |
SX-500BH |
ASX-500 |
|
|
|
|
220V
model
|
230V
model
|
240V
model
|
230V
model
(220 to 240V)
|
120V model
230V model (220 to 240V)
|
230V model
(220 to 240V)
|
120V model
230V model
(220 to 240V)
|
120V model
230V model
(220 to 240V)
|
Operating
Temperature Range |
Sterilizing |
105 to 123℃ *1 |
105 to 135℃ *1 |
105 to 135℃ *1 |
105 to 132℃ |
Heating |
– |
45 to 104℃ |
45 to 104℃ |
– |
Warming |
– |
45 to 95℃ *1 |
45 to 95℃ *1 |
– |
Maximum Operating Pressure |
0.193 MPa |
0.250 MPa |
0.263 MPa |
0.250 MPa |
0.263 MPa |
0.263 Mpa |
Time Setting
Range |
Sterilizing |
0:01 to 9:59 / 1 to 99 hours /
1 to 999 minutes / 0:01 to 9:59,
10 to 99 hours (Selectable) |
0:01 to 9:59 / 1 to 99 hours /
1 to 999 minutes / 0:01 to 9:59,
10 to 99 hours (Selectable) |
0:01 to 9:59 /
1 to 99 hours /
1 to 999 minutes /
0:01 to 9:59,
10 to 99 hours (Selectable) |
0:01 to 9:59 |
Heating |
– |
– |
Warming |
– |
1 to 99 hours |
1 to 99 hours |
– |
Display |
Temperature |
Digital |
Digital |
Digital |
Digital |
Pressure
Gauge |
Analog |
Analog |
Analog |
Analog |
0 to 0.4 MPa |
0 to 0.4 MPa |
0 to 0.4 MPa |
0 to 0.4 MPa |
Time |
Digital |
Digital |
Digital |
Digital |
Safety Devices |
Safety Valve
Lid Interlock
Leakage Breaker
No Water Heating Prevention
Over-temperature Prevention
Over-pressure Prevention
Temperature Sensor
Malfunction Detection
Pressure Sensor Malfunction Detection
Exhaust Valve Malfunction Detection
Low Sterilizing Water Level Detection
Chamber Lid Open / Close Detection
Drain Access Cover Open Detection
Exhaust Tank Full Level Detection |
Safety Valve
Lid Interlock
Leakage Breaker
No Water Heating Prevention
Over-temperature Prevention
Over-pressure Prevention
Temperature Sensor
Malfunction Detection
Pressure Sensor Malfunction Detection
Exhaust Valve Malfunction Detection
Low Sterilizing Water Level Detection
Chamber Lid Open / Close Detection
Exhaust Bottle Detection |
Safety Valve
Lid Interlock
Leakage Breaker
No Water Heating
Prevention
Over-temperature
Prevention
Over-pressure
Prevention
Temperature Sensor
Malfunction Detection
Pressure Sensor
Malfunction Detection
Exhaust Valve
Malfunction Detection
Low Sterilizing
Water Level Detection
Chamber Lid
Open / Close Detection
Exhaust Bottle Detection |
Safety Valve
Lid Interlock
Leakage Breaker
No Water Heating
Prevention
Over-temperature
Prevention
Over-pressure
Prevention
Temperature Sensor
Malfunction Detection
Pressure Switch
Malfunction Detection
Exhaust Valve
Malfunction Detection
Low Sterilizing Water
Level Detection
Chamber Lid
Open / Close Detection |
Chamber |
Dimensions
(mm) |
φ 450 × 678 |
φ 370 × 774 |
φ 325 × 733 |
φ 325 × 553 |
φ 370 × 774 |
φ 325 × 733 |
φ 325 × 733 |
Capacity |
100 L |
79 L |
58 L |
44 L |
79 L |
58 L |
58 L |
Material |
SUS304 (Stainless Steel) |
SUS304 (Stainless Steel) |
SUS304 (Stainless Steel) |
SUS304 (Stainless Steel) |
Main Body |
Dimensions
(mm)
(Excluding
protruding) |
600W × 600D × 940H |
470W
×
528D
×
1,003H |
410W
×
477D
×
970H |
410W
×
477D
×
790H |
470W
×
528D
×
1,003H
(Bio Box:
φ215 × 525 H) |
410W
×
477D
×
970H
(Bio Box:
φ215 × 525 H) |
410W × 477D × 970H |
Net Weight |
108 kg |
72 kg |
60 kg |
50 kg |
72 kg
(Bio Box: 5 kg) |
60 kg
(Bio Box: 5 kg) |
60 kg |
Power Requirements |
AC220V
50 / 60Hz
15A 1P |
AC230V
50 / 60Hz
15A 1P |
AC240V
50 / 60Hz
15A 1P |
AC230V
50 / 60Hz
15A 1P |
AC120V
50 / 60Hz
20A 1P
AC230V
50 / 60Hz
15A 1P |
AC120V
50 / 60Hz
15A 1P
AC230V
50 / 60Hz
10A 1P |
AC230V
50 / 60Hz
15A 1P |
AC120V
50 / 60Hz
20A 1P
AC230V
50 / 60Hz
15A 1P |
AC220 to 240V 50 / 60Hz
15A 1P |
Standard Basket |
Stainless Steel Basket
( φ 415 × 236 mm)
2 pcs |
Stainless
Steel
Basket
( φ 345
×
181 mm)
2 pcs |
Stainless
Steel
Basket
( φ 300
×
181 mm)
2 pcs |
Stainless
Steel
Basket
( φ 300
×
181 mm)
1 pc |
Stainless
Steel
Basket
( φ 345
×
181 mm)
2 pcs |
Stainless
Steel
Basket
( φ 300
×
181 mm)
2 pcs |
– |
Compliance of Standard |
IEC 61010-2-040, IEC61326-1 |
– |
– |
– |